Technical Guide - Infield Verification following ASTM E83 -23

This guide provides details on Vector operation at the specific points throughout ASTM E83 -23 Section 5 - Verification Procedure for Extensometer Systems where the process differs from the standard method with traditional extensometers.
Axial and Transverse verifications are performed seperately but use a very similar process. The differences are highlighted below.

Setup

5.3 Position of Extensometer,
5.5 Temperature Control,
5.6 Method of Reading.
  1. Set-up, plug-in, and power-on the Vector.

    Confirm Vector enters scanning, with an oscillating blue light.
    Confirm Vector connects to the Interface App.
  2. Mount the Vector infront of the verification device.

    Ensure the verification device is positioned securely, the Vector is mounted suitably, and the distance between the instruments is within specification.
    For more information, see sections Mounting, Connections, and Installation and Positioning Vector of the User Manual.
    _images/verif_mounted.png

Information - transverse verifications

info
Transverse verifications are performed seperately to axial verifications and involve different spindles and mounting.
Refer to the later section for the transverse procedure.
  1. Allow Vector to acclimatise to the laboratory environment and temperature.

    Ensure Vector is powered ON and in scanning mode for a recommended 20 minutes.
  2. Configure the measurement type and units.

    Measurement and Units are set within the Interface App - Settings:
    _images/ScanningSettings.png
    If extension or elongation is required for verification (for example, deflection ISO 5893), configure the measurement type as extension.
    _images/OptionsMeasurement.png
    If calibrating transverse, set Vector to Biaxial in the Interface App - Settings.
    _images/OptionsTransverse.png
  3. Configure the measurement output.

    Analogue scaling is configured within the Interface App - Settings - Output.
    Confirm the scaling is correct for both the Vector and UTM using the set output options.
    _images/OutputASet.png
    For more information, see Scaling the Analogue Output Signal of the User Manual.
    If calibrating transverse, confirm the scaling is correct for the transverse channel in the same way.
    _images/OutputATransverse.png

Information - digital output

info
When using a digital connection
Refer to the relevant Vector tech guide for configuration.
Scaling is not required when using digital. The analogue scaling has no effect on the digital output.

Information - analogue output

info
When using an analogue connection
Refer to the relevant OEM information from the UTM manufacturer for configuration.
Scale the output to accomodate the desired strain or displacement ranges over which the verification is to be performed.

Inspection

  1. Inspect the optical windows on the front of the Vector for scratches, smears, dust, or debris.
    Note the initial condition is “As Found”.
    If verifications are reattempted, clean the front of Vector with a clean microfibre cloth.
  2. Inspect the condition of the power and data cables for damage, especially the analogue connection.
    Confirm that the recieved analogue signal at the UTM is not affected by movement of the data cables.
    If noise is present, change the cables.
  3. Inspect the reported class in the Interface app when using the validation block provided.
    Confirm the class is B-1 throughout the rated field of view.
    If lower classes are shown consistently in the field of view, contact your supplier.
    _images/valBlock1.png

Gauge length

For optical instruments there is no measurement of gauge length error, only the nominal GL entry is required.
Each time Vector enters measuring, Vector automatically sets the gauge for Axial, and Transverse when required. The precise gauge length is reported in the Interface App.
_images/Measuring.png

Rationale

Vector automatically positions the gauge algorithmically using user preferences and presented markings. The extensometer gauge length can vary slightly between scans, irrespective of the presented markings, based on the instantaneous appearance.
The resultant measurement relates directly to the positioned gauge. The extensometer gauge length equals the nominal gauge length of the verification device. Therefore, the outputted strain and extension correspond directly to the stated gauge length, not to any specific feature on the specimen.
Practically, the gauge length will also vary between measurements for a contacting extensometer. However, for contacting extensometers the true gauge length is not known and so a nominal must be assumed and the variability of the gauge lengths from a nominal must be assessed at verification.
The additional information of a precise gauge length provided in the Interface App is beneficial in producing a test or verification of higher quality.

Marking and positioning verification device spindles

When Calibrating Axial

  1. Mark the spindles.

    Clean, prepare and mark the spindle blocks in a similar way that will be employed for specimen testing.
    For more information, see Specimen Marking of the User Manual.
    _images/verif_spindles_marks.png
  2. Position spindles to represent the specimen.

    Ensure spindles are square and aligned.
    For rings and filled circles, space the spindles accordingly to represent the gauge.
    For speckled spindles, ensure the spindles are mounted closely together to form a single, contiguous speckle region.
    _images/verif_spindles_mounted.png
  1. Mount the Vector infront of the verification device.

    Ensure the verification device is positioned securely, the Vector is mounted suitably, and the distance between the instruments is within specification.
    For more information, see sections Mounting, Connections, and Installation and Positioning Vector of the User Manual.
    _images/verif_mounted.png

When Calibrating Transverse

  1. Mark the spindles.

    For transverse, the verification process is performed with the extensometer perpendicular to the verification device. Specific spindles are required to calibrate the transverse.
    Clean, prepare and mark the spindle blocks in a similar way that will be employed for specimen testing.
    For more information, see Specimen Marking of the User Manual.
    _images/verif_spindles_trans_tape.png
    The example shown is for a 3590VHR, using a standard flat spindle and a custom transverse “paddle”. Alternative custom paddles may be required for other verification devices.
    For convenience, tape the flat spindle and paddle together so that the spindle end and the paddle are almost touching. This is the spindle gap and the centre line of the setup.
    _images/verif_spindles_trans_marks.png
    On the other side, mark the setup with appropriate markings for the verification. As the axial measurement is required but untested, the gauge length is not relevant.
    Ensure that the markings are symmetric about the centre line of the setup; i.e. ensure that the marking length is vertically centred to the spindle gap.
  2. Position spindles to represent the specimen.

    Ensure the paddle is connected to the fixed head of the verification device.
    Ensure spindles are square and aligned.
    For rings and filled circles, space the spindles accordingly to represent the gauge.
    For speckled spindles, ensure the spindles are mounted closely together to form a single, contiguous speckle region.
    _images/verif_spindles_trans_mounted.png
  3. Mount the Vector infront of the verification device.

    Ensure the verification device is positioned securely, the Vector is mounted suitably, and the distance between the instruments is within specification.
    Position Vector horizontally, either by using the dedicated “landscape orientation intermediary” or by rotating the extension arm extrusion a quarter turn and reinserting into the extension arm housing.
    For more information, see sections Mounting, Connections, and Installation and Positioning Vector of the User Manual.
    _images/verif_trans_laser.png
    Ensure that the central laser passes through the spindle gap. Ensure that the bottom laser is just outside the edge of the paddle, such that both axial and transverse measurements are performed.

Information - transverse verifications

info
Ensure the paddle spindle is attached to the fixed head of the verification device

Verification process

  1. Position the Vector at the correct distance from the verification device.

    Adjust, measure, and record the distance between the front face of Vector and the front face of the verification device blocks. Refer to the user manual for operating limits.
    _images/verif_measure_distance.png
    Alternatively to a ruler, the Vector reports the distance to markings and validation block within the Vector Interface App.
    _images/Scanning.png _images/Validating.png
  2. Align the Vector to the spindles using the laser alignment guides.

    Ensure:
    • The central laser aligns with the central travel of the verification device.

    • The lower laser is just below the markings.

    Ensure the mounting is secured and the locking levers are firm. For more information, see section Positioning Vector of the User Manual.
    _images/verif_axial_laser.png
  3. Prevent any Obstructions or Vibrations.

    Ensure the Vector is not obstructed during a verification, including any blockage to the lights.
    Ensure the Vector and verification device are securely mounted and in a still environment.
    If required, allow the system to settle for a second after each motion.

Preperation Before Each Verification Run

With both the verification device and Vector positioned and configured as intended, ensure the three steps below are followed.
  1. Zero Vector Before Each Run

    Reset Vector with a “handstop”.
    Cover the markings for atleast 1 second to reset Vector.
    _images/handstop.png
    The status lights on Vector will go from green to oscilliting blue to indicate Vector has stopped measuring.
    Alternatively, the “hand-stop” (Vector Reset) button in the interface can be used.
    _images/MeasuringHandstop.png
  2. Record the Initial Gauge Length

    Ensure Vector is in measuring mode, signified by a solid green light.
    Ensure the gauge length is as intended.
    When using speckles, ensure the target gauge length is specified via Options or using the Edit on the Scanning page.
    _images/OptionsTGL.png _images/ScanningTGL.png
    Refer to the Vector Interface dashboard to help diagnose issues.
    For each run, record the initial gauge length precisely as that listed by the Vector Interface.
    _images/Measuring.png
  3. Take Readings from the System Output
    When using an analogue connection, take readings of the strain and extension from the UTM and test control software.
    When using a digital connection, the Interface App can be used if more convenient as the values will be identical.
    _images/verif_ReadFromSystem.png

Resolution

In all cases, with analogue or digital, take readings from the system output.
The displayed value of strain or extension within the Interface App is not live, raw data.
_images/verif_ReadFromSystem.png