distance Measurement Ended Outside of Operating Distance Range

The Vector was positioned where the detected markings moved either too far from, or too near to Vector.
The permitted Operating Distance range extends 250mm to 350mm from Vector.
_images/operating.png

Considerations

If Measurement ended unexpectedly and this indicator is active, consider the following causes:

  1. The Measurement started with Vector close to the limit of the Operating Distance range.
    Check the Vector is positioned correctly.
    See section on Vector Outside of Operating Distance for more information.
  2. The load-train moved the specimen towards or away from Vector during the start of the test.
    Check that the arrangement of grips and fixtures do not allow excessive travel and that the load train is tight before Measuring.

See Vector Outside of Operating Distance for more information.

Information

info

The Operating Distance is measured from the front face of the Vector.